33
8 Digital Communication Interface
Interface Commands
Command Description
Command
Format
†
Command
Example
‡
Returned
Returned
Example
Acceptable
Values
Calibration Parameters Menu
Read R0 calibration parameter r[0] r r0: 999.999 r0: 100.578
Set R0 calibration parameter to
n
r[0]=n r=100.324 100 to 105
Read ALPHA calibration
parameter
al[pha] al al: 9.9999999 al: 0.0038573
Set ALPHA calibration parameter
to
n
al[pha]=n al=0.0038433 0.002 to 0.006
Read DELTA calibration
parameter
de[lta] de de: 9.9999 de: 1.507
Set DELTA calibration parameter de[lta]=n de=1.3742 0.5 to 1.9
Read BETA calibration parameter be[ta] be Be: 9.999 be:0.342 —25 to 25
Set BETA calibration parameter be[ta]=n be=0.342
Miscellaneous
Read firmware version number *ver[sion] *ver ver.9999,9.99 ver.9009,1.21
Read structure of all commands h[elp] h list of commands
Read ALL operating parameters all all list of parameters
Legend: [] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note: When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
†
Preface all commands with either C: (for cold side) or H: (for hot side). If the C: or H: is left off, the returned value is for the hot side.
‡
The returned command has either a ‘C’ or an ‘H’ before the colon depending on which side was queried.
9009 Communication Commands Continued